IOM

Neuro-TES

Transcranial Electrical Stimulator 

usb USB Bağlantı
iso ISO Belgelendirme
dotnet Yazılım .NET Platformunda

Neuro-TES electrical stimulator is intended for the diagnostic transcranial stimulation of motor cortex to study the motor evoked potentials at intraoperative monitoring.

Neuro-TES Operation

The electrical stimulator can be used both as a part of Neuro-IOM system and in stand-alone mode (under Neuro-TES software control). In stand-alone mode the stimulator can be synchronized with third-party recording system.

 

 

Main Specifications

Stimulus amplitude — 0–1000 V, 5 V step

Pulse duration — 40 μs, 50 μs, 60 μs, 70 μs, 80 μs

Stimulus waveform — rectangular monophasic

Number of stimuli in train — 1–9

 

 

 

 

Anode and Cathode Switch

The stimulator has a built-in electronic switch, which allows switching anode and cathode of electrical stimulator to any outputs of junction unit. Junction unit can be located in the patient zone at the distance of up to 5 m from main unit and control PC. The commutation is performed from Neuro-TES software.

The electrical stimulator can be supplied with mechanical switch as well. In this case, anode and cathode switch is performed using knobs on the front panel.

 

 
neurotes1Neuro-TES software
neurotes2TES Electronic Switch
neurotes3The mechanical switch and junction unit
  • Neuro-TES transcranial electrical stimulator control unit
  • Cable for stimulating electrode, touch-proof
  • Neuro-TES software
  • Technical manual

 

In 2011, I purchased the first Evidence 3102* from Schreiber&Tholen**, Germany, for my research lab dealing with disease models of peripheral neuropathies, myasthenic syndromes, and of myelopathies in mice and rats. The run-in period was very short because the programs are user-friendly, dependable and adaptable to the special needs in small rodents, the youngest being 8 days old. No EMG machine I ever used before was as powerful as this one. On numerous trips to collaborating labs in the US, Germany and Switzerland I carried this equipment with me. Meanwhile 6 completely established sets are working in most of these collaborating labs. There was no failure ever of the Neurosoft hardware, and in case of minor software problems Schreiber&Tholen provided swift and concise help through the Internet around the globe if needed — again with no precedence in my career in clinical electrophysiology. Even in lab facilities without perfect electromagnetic shielding little if any problems occurred. I highly recommend this equipment to anyone interested in doing clinical neurophysiology in animal models of neurologic disorders

*  Evidence 3102 — Neuro-MEP-Micro 2-channel Ultraportable EMG and NCS System with the Evidence trademark.
**Schreiber&Tholen —  the official distributor of Neurosoft company in Germany, Austria, Switzerland.<

avmlogo

AVM Sağlık Sistemleri ürün ve hizmetleri konusunda ilgileniyor ve yeniliklerden haberdar olmak istiyorsanız; bizi takip edin...

encoksatan logo

orjinal-urun

İletişim

AVM Sağlık Hizmetleri Ltd. Şti.

Küçükbakkalköy Mh. Gül Ağacı Sk.
No:13 34750 Ataşehir - İstanbul

Tel: +90 (216) 368 18 00
Faks: +90 (216) 368 19 00

E-posta: bilgi@avmsaglik.com.tr